A device history record should let a reviewer trace a released device or batch back to the approved configuration, materials, manufacturing steps, inspection results and release decision. For an OEM outsourcing PCBA, the difficult work is often connecting records held by different organizations. A folder full of reports can still leave a gap between the boards shipped and the evidence reviewed.
FDA’s Quality Management System Regulation became effective on February 2, 2026 and incorporates ISO 13485:2016. The familiar DHR name therefore needs a current requirements map. This guide proposes a practical record handoff between a device OEM and an assembly supplier. The manufacturer responsible for the finished device should determine the applicable regulatory requirements, product scope and release authority.
What changed for the device history record under QMSR?
FDA explains in its final rule, response to Comment 31, that QMSR does not retain separate requirements for the old DHR, DMR and DHF record types. Their relevant elements are largely addressed through ISO 13485 documentation requirements. FDA specifically connects many former DHR elements with the medical device or batch record described in clause 7.5.1.
A company can assess its existing records against that framework without assuming a new folder name completes the transition. Keep a mapping that identifies the requirement, the record supplying evidence, its owner and the retrieval location. If an old form captures useful production history, assess its contents and controls before replacing it. A changed title does not repair a missing inspection result.
Separate the approved manufacturing instructions from the history of a particular build. One describes what should happen; the other records what happened to identified units. The history should point to the instruction version used at the time. Linking every older lot to today’s procedure would erase the distinction a later reviewer needs.
For supplier onboarding, assign someone to review the cross-reference with the OEM’s quality and regulatory team. Ask them to identify gaps that affect acceptance, traceability or record access. Keep the completed mapping with its approval and revision. The supplier can propose a practical export format while the responsible manufacturer decides whether that format meets its obligations.
Set the product and batch boundary before collecting files
Define what a device history record covers in this relationship. An assembly supplier may ship a populated PCB, a programmed module or a tested subassembly. Those items can enter a larger device manufactured elsewhere. Record the assembly identity and handoff point so an internal board record cannot be mistaken for the entire finished-device history.
Agree on the identifiers before the pilot lot. State how the OEM product number maps to the supplier assembly number, PCB revision, BOM revision, firmware version and manufacturing variant. Identify which fields are printed on a board and which remain in the system. The mapping should survive a label replacement or a later transfer between facilities.
Batch splits need explicit links. If one incoming lot feeds two work orders, both outputs should remain connected to their input. When several subassembly lots enter one finished-device batch, retain each relationship. A single free-text comment may be hard to search when an investigation begins with a component lot rather than a finished serial number.
Verify PCBA traceability before the purchase order by asking for a bounded retrieval demonstration. Use a redacted example with the same proposed identifier structure. A generic presentation about an MES cannot show whether the chosen fields connect across the actual supplier and OEM systems.
Agree what each quantity means: started, completed, held, scrapped, returned or transferred. Repeated testing must not create a second finished unit in the count. Likewise, a duplicate export should not appear as additional production. Use a consistent unit identifier and event relationship when reconciling these records.
If quantities cannot be reconciled, identify the missing transaction before signing the handoff. The record should explain the status of affected units rather than hide the difference in a spreadsheet adjustment.
Link each carrier position to an identified lot or unit before combining production records.
Map device history record inputs to build history
Start the record index with the released manufacturing configuration. The file identities should resolve to the versions authorized for that work order, including any approved deviations. Document control should be able to answer which change became effective at which unit, lot or operation.
Use the BOM revision effectivity record when a material or population change intersects production. The production history should preserve the configuration used before the change as well as the configuration used afterward. Keep the rationale and authorization for any units accepted under an exception.
This table is a proposed index, not an exhaustive regulatory checklist. Add product-specific records when the manufacturing route requires them. Conversely, identify why a field is inapplicable instead of filling it with a misleading zero, a copied approval or an unrelated attachment. A reviewer should be able to distinguish missing evidence from an operation that the approved route never required.
Keep material and process evidence tied to identified units
Material traceability begins with identity, not with a photograph of a labeled reel. Decide which lot, date-code, receiving and supplier fields the project needs. Preserve the route from received material through issue, consumption, return and replacement. If a reel changes during a work order, define how the transition connects to the affected boards.
Record controlled substitutions where they occur. A purchasing note allowing a different supplier does not necessarily authorize a different manufacturer part number. The approved source hierarchy should settle that distinction. Link the decision to the actual material used, the affected quantity and the revision or deviation that permitted it.
Process evidence should identify the execution that matters to the product. Examples include the program loaded for a machine operation, a required inspection status, or the version of a work instruction. Record the applicable method and result rather than collecting every machine screenshot indiscriminately. Large exports are difficult to use when no field identifies the work order or operation.
For an illustrative split lot, imagine that a connector substitution becomes effective after the first portion of an order is built. The history needs an identifiable boundary and evidence for both configurations. An undated note saying “alternate approved” cannot establish which boards contain the alternate. Resolve that gap before the affected units leave the agreed hold point.
Preserve parent-panel and individual-unit relationships when required.
Retain failures in the device history record
A device history record should preserve the path to the final disposition. Define how first-pass failures, repeated measurements, repair or rework actions and later results remain associated with the same unit. Keeping only the last green result can hide a sequence the OEM needs when investigating an intermittent fault.
For each relevant test, retain the identity of the tested object, the applicable method or program, the conditions needed to interpret the measurement, and the recorded result. A pass flag may be sufficient for some agreed checks; other checks need measured values or supporting files. Decide that distinction with the requirement owner before selecting an export format.
Document an authorized rework as an event with its own scope and evidence. Name the affected references, method, operator, replacement material where relevant and verification after intervention. Link any change in design intent to the PCBA engineering change-control process . A revised board and an unchanged board should not acquire identical histories through a bulk spreadsheet update.
Keep unresolved findings visible during review. Record whether they prevent supplier shipment, OEM acceptance or a later manufacturing step. Each condition needs an owner and the evidence required to close it. A signed certificate of conformance should not silently override an open condition in the agreed release package.
Record the configuration at each controlled process transfer.
Define record ownership and electronic handoff
Agree who creates, reviews, retains and supplies each record. The supplier’s internal database and the OEM’s quality system may use different structures. A useful handoff includes a field map, file naming rules, record versions and a way to identify missing or replaced exports. Specify the authorized request path for information that cannot be sent routinely.
For electronic records, ask the responsible quality and regulatory team to assess applicable requirements, including FDA’s Part 11 scope and application guidance. A spreadsheet, PDF export or software feature list alone does not demonstrate that the particular record workflow has been assessed and controlled. Test retrieval, review, corrections and export using the intended process.
Under 21 CFR 820.35(c), UDI recording is an additional requirement for each medical device or batch of medical devices. Have the OEM define the applicable device identification and its link to internal assembly identifiers. An EMS board serial number should not be represented as the finished device’s UDI without that defined relationship.
Use the MES inspection and test data structure to support the exchange, then verify the exported package independently. Check permissions, legibility, timestamps and links after transfer. A record available to a supplier administrator may still be inaccessible to the OEM reviewer who must make the decision.
Test the device history record before authorizing release
Choose a representative unit or batch and ask a reviewer who did not assemble the package to reconstruct its history. Start with the shipment or subassembly identifier. Find the governing configuration, material links, required operations, test results, exceptions and authorized release. If the reviewer needs an oral explanation to bridge a gap, record the missing link and its owner. Keep the affected acceptance decision open until the controlled package is corrected and the retrieval check succeeds.
Test retrieval in the opposite direction too. Starting from a selected material lot, determine which assemblies could be affected. The requested scope should be explicit; finding one matching unit does not prove the complete population has been located. Retain the query basis and the reviewed result so another person can repeat the exercise.
Agree on retention and access obligations for the product, market and contract rather than adopting a convenient default period. Include supplier changes, system migration and the end of the commercial relationship in that discussion. Close the handoff only when the responsible reviewer can retrieve the agreed evidence and explain the release decision from the record.
Physical segregation supports a record system but cannot replace it.
Device History Record FAQ
Can we continue calling our production package a device history record?
The name can remain useful internally. Map its contents and controls to current applicable requirements, and document any gaps. A familiar form name does not establish completeness under QMSR.
Does one PCBA lot record cover the finished medical device?
Only the defined assembly scope. The OEM should connect it to other manufacturing, labeling, acceptance and device-identification records required for the finished product. Define that boundary in the supplier agreement.
Must the supplier email every raw machine file with every shipment?
Set the delivery and retrieval requirements from the product risk, accepted process and applicable obligations. Identify which files accompany shipment and which remain controlled and retrievable through an agreed request path.
What should happen when a historical record contains an error?
Use the approved correction process. Preserve the original entry or history where required, identify the correction and its authority, and review whether the error changes the affected unit or batch disposition.
To define a PCBA record handoff, send your assembly identity rules, required record fields, revision controls and proposed release responsibilities through the GNS contact form . Ask for a documented evidence-delivery scope and a sample retrieval check before approving the production package.